Simplicity in design
- Two button operation
- Clean interface
- Robust & Waterproof IP65
- Size of a mobile telephone
- Cyber secure
- GPS positioning (switchable on/off)
- Includes carry case
Easy to maintain
- Self tests daily, weekly and monthly
- Pads 5 years / LiH battery up to 3 years
- Choice of Rechargeable or disposable battery
- Carry case included
Integrity
- System integrity test on battery insertion
- Detects missing pads
Patient analysis system
- Automatically evaluates patient impendence for proper pads contact
- Monitors signal quality and analysis for patient ECG for shockable/non-shockable rhythms
- Meets AHA/ERC recommendations
Physical parameters
Size (height*width *length)
- 2.09in*3.39in*5.91 in
- 5.3cm*8.6cm*15cm
Weight (including battery)
- 1.54 lb. (0.7kg)
Tolerable impact/falling damage
- Free to fall from a height of 1.5 m on a hard surface
Environmental parameters
Operation condition
- Temp: -15°C to 50°C (Can work for at least 60 minutes after moving from room temperature to -20°C)
- Relative humidity: 0% to 95%, non-condensing
- Atmospheric pressure range: 50.4kPa to 106kPa
Storage condition
- Short-term Temp: -40°C to 70°C (Within one week)
- Long-term Temp: 0°C to 50°C
- Relative humidity: 0% to 95%, non-condensing
- Atmospheric pressure range: 59.4kPa to 106kPa
Transportation condition
- Short-term Temp: -40°C to +70°C (Within one week)
- Long-term Temp: 5°C to 50°C
- Relative humidity: 0% to 95%, non-condensing
- Atmospheric pressure range: 59.4kPa to 106kPa
Altitude
- 4500m
Defibrillation
Shockable Rhythm
- VT/VF
Waveform
- Truncated biphasic exponential waveform
Energy level
- Automatic pre-programmed selection
- (Adult mode: 150J; Pediatric mode: 50J)
Output control
- Manual operation (In rescue mode, the shock button should be pressed manually).
Operational impedance limitation of patient
- 200 to1800